9 Position Heating/Cooling Sample Block (9CB)

9 position peltier-driven heating-cooling sample changer for SANS

A commercially available peltier driven temperature block is available at the CHRNS 30-m SANS. The block was designed and manufactured by Quantum Northwest. The 9 position block has 3 independently controlled temperature zones, each controlling 3 sample wells. Standard operations allow a temperature range of -10 °C to 110 °C, with a total temperature range of -40 °C to 140 °C. Reaching  temperatures outside the standard range requires adjustment of the sink temperature.

The main advantage of the block over others is the time to equilibration. Fluid driven blocks can take up to an hour for the sample temperature to reach the desired temperature. The 9CB, operating under standard conditions, is rapid enough to equilibrate the sample temperature in 10 minutes or less for temperature changes of 50 degrees or less.

To ensure optimal temperature stability, quartz windows are installed before and after the sample. Quartz is optically transparent and transmits a laser beam used for the alignment of the sample block.

Sample changers with 9 positions are currently available at the CHRNS 30m SANS, and CHRNS VSANS instruments. Another 9-position changer has arrived for use at the nSoft 10m instrument, and will be available shortly. A fourth block for use on NG7 SANS should be available by the end of 2019.