This 7T vertical field wet magnet system has no persistent switch. There are two options for inserts: a top loading CCR insert (4K-600K) and a bottom loading 3He insert.
| Magnet Properties |
|---|
| Max Field | 7 T | | Current at 7 T | 89.97 A | | Field Calibration | 0.0778 T/Amp | | Magnet Resistance at 77K | 80.4 ohms |
| | Magnet Cool Down Info |
|---|
| LN2 Pre-Cool Time | 1 Hr. | | LHe Cool Time | 1 Hr. | | LHe Consumption | ~70 L. |
|
| Insert Comparison |
|---|
| Insert | Temp. Range | Sample Space Atmosphere | Prep. Time (to 4K in magnet) | Sample Change (Time to 4K) | Drawing | | TL-CCR | 4 - 600K | Helium Gas | 6 hours | 2 hour | Drawing | | 3He insert | 300mK - 300K | Vacuum (below 4K & above 20K) | 2 hours | 4 hours | Drawing |
|
More Information: