Search Publications by: Jeffrey T. Cessna (Fed)

Search Title, Abstract, Conference, Citation, Keyword or Author
Displaying 1 - 25 of 55

Liquid scintillation efficiencies, gamma-ray emission intensities, and half-life for Gd-153

November 9, 2023
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Gulakhshan Hamad, Lizbeth Laureano-Perez, Leticia Pibida, Brian E. Zimmerman
Gadolinium-153 was standardized for activity by live-timed anticoincidence counting and an ampoule was submitted to the international reference system (SIR). Absolute emission intensities for the main γ rays were determined with calibrated high-purity

Primary standardization of 212Pb activity by liquid scintillation counting

September 22, 2022
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Lizbeth Laureano-Perez, Leticia Pibida, Brian E. Zimmerman
An activity standard for 212Pb in equilibrium with its progeny was realized, based on triple-to-double coincidence ratio (TDCR) liquid scintillation (LS) counting. A Monte Carlo-based approach to estimating uncertainties due to nuclear decay data

BIPM comparison BIPM.RI(II)-K1.Gd-153 of activity measurements of the radionuclide 153Gd to include the 2020 result of the NIST (United States)

November 25, 2021
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Lizbeth Laureano-Perez, Leticia Pibida, Carine Michotte, S Courte, M Nonis, R Coulon, S Judge
Since 1988, 3 laboratories have submitted 5 samples of 153Gd to the International Reference System (SIR) for activity comparison at the Bureau International des Poids et Mesures (BIPM), with comparison identifier BIPM.RI(II)- K1.Gd-153. Recently, the NIST

Ra-224 activity, half-life, and 241 keV gamma ray absolute emission intensity: a NIST-NPL bilateral comparison

December 31, 2020
Author(s)
Denis E. Bergeron, Sean Collins, Leticia Pibida, Jeffrey T. Cessna, Ryan P. Fitzgerald, Brian E. Zimmerman, Peter Ivanov, John Keightley, Elisa Napoli
The national metrology institutes for the United Kingdom and the United States have compared activity standards for 224Ra, an -particle emitter of interest as the basis for therapeutic radiopharmaceuticals. Solutions of 224RaCl2 were assayed by absolute

Ohms Law Low-current Calibration System for Ionization Chambers

August 24, 2020
Author(s)
Dean G. Jarrett, Shamith U. Payagala, Ryan P. Fitzgerald, Denis E. Bergeron, Jeffrey T. Cessna, Charles J. Waduwarage Perera, Neil M. Zimmerman
A system for the calibration of electrometers that measure currents from ionization chambers is described. The calibration system uses a 1 GΩ standard resistor in series with a stable voltage source to generate calibration currents from 1 pA to 20 nA

Primary standardization of Ra-224 activity by liquid scintillation counting

October 11, 2019
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Brian E. Zimmerman, Lizbeth Laureano-Perez, Ronald Colle, Leticia S. Pibida, Elisa Napoli
A standard for 224Ra activity has been developed, based on triple-to-double coincidence ratio (TDCR) liquid scintillation (LS) counting. The standard was confirmed by efficiency tracing and 4παβ(LS)-γ(NaI(Tl)) anticoincidence counting, as well as by 4πγ

Standardization of I-124 by three liquid scintillation-based methods

August 9, 2019
Author(s)
Denis E. Bergeron, Leticia S. Pibida, Brian E. Zimmerman, Jeffrey T. Cessna, Ryan P. Fitzgerald
A solution of 124I was standardized for activity by 4πβ(LS)-γ(NaI) live-timed anticoincidence (LTAC) counting, with confirmatory measurements by triple-to-double coincidence ratio (TDCR) and CIEMAT-NIST efficiency tracing (CNET) liquid scintillation

A Reference Material for Evaluation of Cs-137 Radiochronometric Measurements

August 16, 2018
Author(s)
Kevin B. Lavelle, Richard Essex, Jeffrey T. Cessna, Kevin P. Carney, Cole R. Hexel
A new nuclear forensic reference material has been characterized as a standard for radiochronometric determination of the model purification date for 137Cs sources. The purification date of a radioactive source is a potentially diagnostic nuclear forensic

Standardization of 64Cu activity

May 25, 2018
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Leticia S. Pibida, Brian E. Zimmerman
The complex decay scheme that makes 64Cu promising as both an imaging and therapeutic agent in medicine also makes the absolute measurement of its activity challenging. The National Institute of Standards and Technology (NIST) has completed a primary

Activity measurements of the radionuclides 18F and 64Cu for the NIST, USA in the ongoing comparisons BIPM.RI(II)-K4.F-18 and BIPM.RI(II)-K4.Cu-64

January 1, 2018
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Ryan P. Fitzgerald, Leticia S. Pibida, Brian E. Zimmerman, Carine Michotte, Manuel Nonis, Andrew Fenwick, Kelly Ferreira, John Keightley
In 2016, comparisons of activity measurements of 18F and 64Cu using the Transfer Instrument of the International Reference System (SIRTI) took place at the National Institute of Standards and Technology (NIST, USA). This is the first SIRTI comparison for

Results of an International Comparison of Activity Measurements of 68Ge

November 7, 2017
Author(s)
Jeffrey T. Cessna, Ryan P. Fitzgerald, Brian E. Zimmerman, Lizbeth Laureano-Perez, Denis E. Bergeron
An international key comparison, identifier CCRI(II)-K2.Ge-68, has been performed. The National Institute of Standards and Technology (NIST) served as the pilot laboratory, distributing aliquots of a 68Ge/68Ga solution. Results for the activity

Two determinations of the Ge-68 half-life

October 19, 2017
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Brian E. Zimmerman
In nuclear medicine, 68Ge is used to generate 68Ga for imaging by positron emission tomography (PET) and sealed sources containing 68Ge/68Ga in equilibrium have been adopted as long-lived calibration surrogates for the more common PET nuclide, 18F. We

Long-term stability of carrier-added Ge-68 standardized solutions

March 1, 2016
Author(s)
Brian E. Zimmerman, Denis E. Bergeron, Ryan P. Fitzgerald, Jeffrey T. Cessna
Tests for chemical stability were carried out on carrier-added 68Ge solutions prepared and calibrated in 2007 and 2011 to evaluate the suitability of the specific composition as a potential Standard Reference Material. Massic count rates of the stored

Revision of the NIST Standard for 223Ra: New Measurements and Review of 2008 Data

July 1, 2015
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Leticia S. Pibida
After discovering a discrepancy in the transfer standard currently being disseminated by the National Institute of Standards and Technology (NIST), we have performed a new primary standardization of the alpha-emitter 223Ra using Live-timed Anticoincidence

Determination of photon emission probabilities for the main gamma-rays of 223Ra in equilibrium with its progeny

March 11, 2015
Author(s)
Leticia S. Pibida, Brian E. Zimmerman, Ryan P. Fitzgerald, Lynne E. King, Jeffrey T. Cessna, Denis E. Bergeron
The currently published 223Ra gamma-ray emission probabilities display a wide variation in the values depending on the source of the data. The National Institute of Standards and Technology performed activity measurements on a 223Ra solution that was used

Secondary standards for 223Ra revised

March 5, 2015
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Brian E. Zimmerman
Dose calibrator dial settings reported by NIST in 2010 (ARI v. 68, p. 1367) are now known to give erroneously low (by 10 %) activity readings. The original determinations were based on a chain of calibrations; a broken link in this chain was recently

A review of NIST primary activity standards for 18F: 1982 to 2013

August 27, 2014
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Bert M. Coursey, Ryan P. Fitzgerald, Brian E. Zimmerman
The new NIST activity standardization for 18F differs from results obtained between 1998 and 2008 by 4 %. The new results are considered to be very reliable; they are based on a battery of robust primary measurement techniques and bring the NIST standard

Dose calibrator manufacturer-dependent bias in assays of 123I

March 26, 2014
Author(s)
Denis E. Bergeron, Jeffrey T. Cessna, Daniel B. Golas, Rheannan K. Young, Brian E. Zimmerman
Calibration factors for commercial ionization chambers (i.e. dose calibrators) were determined for a solution of 123I; the activity was based on the 1976 NBS standard. A link between the NIST standard and the International Reference System (SIR) was