Skip to main content

NOTICE: Due to a lapse in annual appropriations, most of this website is not being updated. Learn more.

Form submissions will still be accepted but will not receive responses at this time. Sections of this site for programs using non-appropriated funds (such as NVLAP) or those that are excepted from the shutdown (such as CHIPS and NVD) will continue to be updated.

U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Spins Instrument Details

 
Guide NG5,50 mm (width) × 120 mm (height)
Monochromator PG(002) with variable vertical focusing
  Ei from 2.4 to 14 meV
  wavelengths from 2.42 - 5.84 Å
Flux at Sample Position 3.9 × 106 neutrons/cm2/s at 4 Å and open/40' collimation
  7.6 × 106 neutrons/cm2/s at 4 Å and open collimation
Sample Scattering Angles -4° ≤ 2θs ≤ 122°, with additional restrictions depending on Ei
Analyzer 11 PG(002) blades for horizontal focusing or flat modes
  mosaic of individual blade ~30′ FWHM
Collimators source-monochromator - open guide (60/ki)
  monochromator-sample - 10′, 20′, 40′, or 80′
  sample-analyzer - 10′, 20′, 40′, 80′, or radial
  analyzer-detector - 20′, 40′, 80′, or radial
Beam Filters before sample: PG or cooled Be
  after sample: cooled Be or BeO
Detectors: 3He single detector (2.54 cm (diameter) × 15.09 cm (height)
  3He single detector (5.08 cm (diameter) × 15.09 cm (height)
  3He position sensitive detector (20 × 25 cm2 area)
Created April 3, 2019
Was this page helpful?