Spins Instrument Details

 
Guide NG5,50 mm (width) × 120 mm (height)
Monochromator PG(002) with variable vertical focusing
  Ei from 2.4 to 14 meV
  wavelengths from 2.42 - 5.84 Å
Flux at Sample Position 3.9 × 106 neutrons/cm2/s at 4 Å and open/40' collimation
  7.6 × 106 neutrons/cm2/s at 4 Å and open collimation
Sample Scattering Angles -4° ≤ 2θs ≤ 122°, with additional restrictions depending on Ei
Analyzer 11 PG(002) blades for horizontal focusing or flat modes
  mosaic of individual blade ~30′ FWHM
Collimators source-monochromator - open guide (60/ki)
  monochromator-sample - 10′, 20′, 40′, or 80′
  sample-analyzer - 10′, 20′, 40′, 80′, or radial
  analyzer-detector - 20′, 40′, 80′, or radial
Beam Filters before sample: PG or cooled Be
  after sample: cooled Be or BeO
Detectors: 3He single detector (2.54 cm (diameter) × 15.09 cm (height)
  3He single detector (5.08 cm (diameter) × 15.09 cm (height)
  3He position sensitive detector (20 × 25 cm2 area)