Nanoscale Specific Heat Capacity Measurements Using Optoelectronic Bilayer Microcantilevers
Published
Author(s)
Brian G. Burke, William Alexander Osborn, Richard S. Gates, David A. LaVan
Abstract
We describe a new technique for optically and electrically detecting and heating bilayer microcantilevers (Pt−SiNx) to high temperatures at fast heating rates for nanoscale specific heat capacity measurements. The bilayer microcantilever acts simultaneously as a heater, temperature sensor, and mass sensor (0.01 ng resolution). The calibration of the system was validated for a mass of 0.72 ng by melting point and specific heat capacity measurements of deposited aluminum nanoparticles.
Burke, B.
, Osborn, W.
, Gates, R.
and LaVan, D.
(2012),
Nanoscale Specific Heat Capacity Measurements Using Optoelectronic Bilayer Microcantilevers, Nature Methods, [online], https://doi.org/10.1063/1.4772477