Pulse Parameter Dependence on Transition Position and Epoch Duration

Published

Author(s)

Donald R. Larson, Nicholas Paulter, David I. Bergman

Abstract

The pulse parameters of amplitudes and transition duration (10% to 90%) are examined as a function of the epoch and position of the transition edge in the epoch. The primary explanations for the observed variations are the pulse aberrations and the finite settling time of both the pulse generator and the measurement instrument (sampling oscilloscope).
Conference Dates
May 18-20, 2004
Conference Location
Como, 1, IT
Conference Title
Proceedings of the 21st IEEE Instrumentation and Measurement Technology Conference

Keywords

epoch, gigahertz, pulse generator, pulse parameter, sampling oscilloscope, setting time, transition duration