Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Polyelectrolyte Multilayer-Treated Electrodes for Real-Time Electronic Sensing Cell Proliferation

Published

Author(s)

Geraldine I. Mijares, Darwin Reyes-Hernandez, Jon Geist, Michael Gaitan

Abstract

The feasibility of using and the performance of non-biological polyelectrolyte multilayer (PEM) films to facilitate cell attachment on titanium-tungsten alloy/gold (TiW/Au) electrodes for the detection of cellular responses in cell assays are investigated. PEMs are used rather than extracellular matrix (ECM) proteins, which are commonly utilized as cell adhesion promoters on planar electrodes, because the integrity of ECM proteins degrades over time and therefore, decreases the lifetime of biosensors. Poly(ethylene imine) (PEI), sodium poly(styrene sulfonate) (PSS), and poly(allylamine hydrochloride) (PAH) were the polyelectrolytes deposited onto the electrode surface to provide NIH-3T3 mouse fibroblast cells with a biocompatible interface. The deposition was verified with atomic force microscopy (AFM) and Fourier transform infrared spectroscopy (FTIR). Cellular response on the PEMs-modified surface was assessed by cell viability measurements and cell morphology imaging. The PEMs-modified electrodes were found to show a 15% increase in impedance due to cell proliferation, and to have an 82% efficiency in impedance detection when compared to an unmodified device. This work demonstrated the use of PEMs-modified electrodes for the continuous monitoring of cell proliferation and for the future application of probing cell confluency in microfluidic cytotoxicity assays.
Citation
Journal of Research (NIST JRES) -
Volume
115
Issue
2

Keywords

biosensors, polyelectrolyte multilayers (PEMs), biocompatible electrodes, cellular impedance, cell confluency detector, cell proliferation monitoring

Citation

Mijares, G. , Reyes-Hernandez, D. , Geist, J. and Gaitan, M. (2010), Polyelectrolyte Multilayer-Treated Electrodes for Real-Time Electronic Sensing Cell Proliferation, Journal of Research (NIST JRES), National Institute of Standards and Technology, Gaithersburg, MD, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=33183 (Accessed December 16, 2024)

Issues

If you have any questions about this publication or are having problems accessing it, please contact reflib@nist.gov.

Created March 31, 2010, Updated October 12, 2021