Construction of a Dual Protease Column, Subzero (-30 °C) Chromatography System and Multi-channel Precision Temperature Controller for Hydrogen-Deuterium Exchange Mass Spectrometry

Published

Author(s)

Jeffrey Hudgens

Abstract

This tutorial provides mechanical drawings, electrical schematics, parts lists, stereolithography (STL) files for producing three-dimensional (3D)-printed parts, initial graphics exchange specification (IGS) files for automated machining, and instructions necessary for construction of a dual protease column, subzero, liquid chromatography system for hydrogen- deuterium exchange mass spectrometry (HDX-MS). Electro-mechanical schematics for construction of two multi-zone temperature controllers that regulate to +/-0.05 °C are also included in this tutorial.
Citation
Journal of Research (NIST JRES) -
Volume
125

Keywords

chromatography, hydrogen-deuterium exchange, mass spectrometry, precision, peptide, protein, proteolysis, proteomics, reference material, temperature control.

Citation

Hudgens, J. (2020), Construction of a Dual Protease Column, Subzero (-30 °C) Chromatography System and Multi-channel Precision Temperature Controller for Hydrogen-Deuterium Exchange Mass Spectrometry, Journal of Research (NIST JRES), National Institute of Standards and Technology, Gaithersburg, MD, [online], https://doi.org/10.6028/jres.125.025 (Accessed October 11, 2024)