Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Computer-Generated Hologram Cavity Interferometry Test for Large X-Ray Mandrels: Design

Published

Author(s)

Guangjun Gao, John P. Lehan, William W. Zhang, Ulf Griesmann, Johannes A. Soons

Abstract

A glancing incidence interferometric test for large x-ray mirror mandrels, using two computer generated holograms (CGH s), is described. The two CGH s are used to form a double pass glancing incidence system. One layout of the CGH-cavity glancing incidence interferometer is studied and factors relating to its first order design are analyzed. A semi-analytical expression for the phase function is derived which avoids high-order ripple errors and improves the efficiency of the CGH pattern generation. A system of coarse and fine fiducials for the alignment of both CGHs and the tested mandrel is discussed. Compared with the traditional methods such as coordinate measuring machines (CMM s) or other scanning metrology systems, the CGH-cavity glancing incidence interferometric test can measure a mandrel surface without mechanical contact, at a high sampling spatial frequency sampling, lower uncertainty, and high speed.
Citation
Optical Engineering
Volume
46
Issue
6

Keywords

CGH, mandrel, x-ray optics, optical testing, interferometry

Citation

Gao, G. , Lehan, J. , Zhang, W. , Griesmann, U. and Soons, J. (2009), Computer-Generated Hologram Cavity Interferometry Test for Large X-Ray Mandrels: Design, Optical Engineering, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=824730 (Accessed March 28, 2024)
Created June 1, 2009, Updated February 19, 2017