COMPARISON OF EXPERIMENTAL TECHNIQUES FOR EVALUATING THE CORRECTION FACTOR OF A RECTANGULAR WAVEGUIDE MICROCALORIMETER

Published

Author(s)

Thomas P. Crowley, Xiaohai Cui

Abstract

We have evaluated the correction factor of a WR-15 (50 to 75 GHz) rectangular waveguide calorimeter using four different techniques. The four methods are in agreement. Our initial uncertainty analysis indicates that the technique with the lowest uncertainty is one that utilized two identical calorimeter core sections back-to-back. This initial uncertainty is smaller than expected which will require us to re-evaluate uncertainty contributions that were neglected.
Proceedings Title
CPEM 2010
Conference Dates
June 14-18, 2010
Conference Location
Daejon

Keywords

microcalorimeter, microwave power, calorimeter

Citation