Aperture Proximity Effects in High Heat Flux Sensors Calibration
Published
Author(s)
A V. Murthy, Benjamin K. Tsai, Robert D. Saunders
Abstract
In transfer calibration of heat flux sensors, a correction for the irradiance distribution across the sensing area may be required when the sensing areas of reference and test sensors are different. A method to calculate this correction using well known equations for the configuration factors is presented. Also, estimates of the correction for test conditions corresponding to the transfer calibration technique in use at the National Institute of Standards and Technology (NIST) are given.
Murthy, A.
, Tsai, B.
and Saunders, R.
(1998),
Aperture Proximity Effects in High Heat Flux Sensors Calibration, Journal of Research (NIST JRES), National Institute of Standards and Technology, Gaithersburg, MD