A potential low magnetic moment standard reference material (SRM) was studied in an interlaboratory comparison. The mean and the standard deviation of the saturation moment ms, the remanent moment mr, and the intrinsic coercivity Hc of nine samples were extracted from hysteresis-loop measurements. Samples were measured by thirteen laboratories using inductive-field loopers, vibrating sample magnetometers, alternating-gradient force magnetometers, and superconducting quantum-interference-device magnetometers. NiFe films on Si substrates had saturation moment measurements reproduced within 5 % variation among the laboratories. The results show
da Silva, F.
, Pappas, D.
and Wang, C.
(2003),
Magnetic Thin Film Interlaboratory Comparison, Journal of Research (NIST JRES), National Institute of Standards and Technology, Gaithersburg, MD