Draft B Report on the Key Comparison CCM.P-K4.2012 in Absolute Pressure from 1 Pa to 10 kPa

Published

Author(s)

Jacob E. Ricker, Jay H. Hendricks, Thomas Bock, Prazak Dominik, Tokihiko Kobata, Jorge Torres, Irina Sadkovskaya

Abstract

This report summarizes the Consultative Committee for Mass (CCM) key comparison CCM.P-K4.2012 for absolute pressure spanning the range of 1 Pa to 10 000 Pa. The comparison was completed via calibration of a transfer standard carried out at six NMIs during the period May 2012 and September 2013. Two nominally identical transfer standard packages were used in the comparison, Package B being circulated among participating laboratories and Package A remaining at the pilot lab in reserve as a back-up unit for the circulating package B. NIST’s experience in the previous low pressure comparison and detailed knowledge in the design and construction of high-stability resonance silicon gauge transfer standard packages [3,4] made it well suited to be the pilot. NIST completed construction and testing of a two new high stability transfer standard packages capable of absolute pressure measurements spanning the extended range of 1 Pa to 10 000 Pa which were used to accomplish the goals of the international key comparison set forth by the BIPM.
Citation
Metrologia

Keywords

Key Comparison, Absolute, Pressure

Citation

Ricker, J. , Hendricks, J. , Bock, T. , Dominik, P. , Kobata, T. , Torres, J. and Sadkovskaya, I. (2016), Draft B Report on the Key Comparison CCM.P-K4.2012 in Absolute Pressure from 1 Pa to 10 kPa, Metrologia (Accessed March 29, 2024)