Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

X-ray computed tomography instrument performance evaluation, Part I: Sensitivity to detector geometry errors

Published

Author(s)

Balasubramanian Muralikrishnan, Katharine M. Shilling, Steven D. Phillips, Wei Ren, Vincent Lee, Felix H. Kim

Abstract

X-ray computed tomography (XCT), long used in medical imaging and defect inspection, is now increasingly used for dimensional measurements of geometrical features in engineering components. With widespread use of XCT instruments, there are growing calls for the development of standardized test procedures to verify manufacturer specifications and provide pathways to establish metrological traceability. As technical committees within ASME and ISO are developing documentary standards that include test procedures that are sensitive to all known error sources, we report on work exploring one set of error sources, instrument geometry errors, and their effect on dimensional measurements. In particular, we have studied detector and rotation stage errors in cone-beam XCT instruments, and determined their influence on sphere center-to- center distance errors and sphere form errors for spheres located in the tomographically reconstructed measurement volume. We have developed a novel method called the single point ray tracing method that allows for efficient determination of the sphere center-to-center distance error and sphere form error in presence of each of the different geometry errors in an XCT instrument. In this first Part, we describe the single point ray tracing method, discuss optimal placement of sphere so that sphere center-to-center distance errors and sphere form errors are sensitive to the different detector geometry errors, and present data validating our method against more conventional radiograph-based tomographic reconstruction method. In Part II of this work, we discuss optimal placement of spheres so that sphere center-to-center distance errors and sphere form errors are sensitive to error sources associated with the rotation stage. This work is in support of ongoing standards development activity within ASME and ISO for XCT performance evaluation.
Citation
Journal of Research (NIST JRES) -
Volume
124

Keywords

cone-beam, distance error, documentary standards, form error, geometry errors, performance evaluation, radiograph-based method, sensitivity analysis, single point ray tracing method, X- ray computed tomography

Citation

Muralikrishnan, B. , Shilling, K. , Phillips, S. , Ren, W. , Lee, V. and Kim, F. (2019), X-ray computed tomography instrument performance evaluation, Part I: Sensitivity to detector geometry errors, Journal of Research (NIST JRES), National Institute of Standards and Technology, Gaithersburg, MD, [online], https://doi.org/10.6028/jres.124.014 (Accessed April 19, 2024)
Created July 10, 2019, Updated March 1, 2021