Lung diseases such as chronic obstructive pulmonary disease are a major health burden to the society, where early detection plays a crucial role for treatment success. For detection, as well as diagnosis and serial evaluation, imaging plays a major role, but lung diseases are often still diagnosed in progressed states for which effective causal therapies do not presently exist. Recently, dark-field lung imaging has been introduced as a promising technique for early stage detection of alterations in lung micro-structures. This work presents an analyzer-free, full-scale lung imaging system based on a dual-phase interferometer, which allows tuning and direct resolution of grating induced intensity fringes. It provides the classical absorption chest image with additional dark-field information without significant attenuation of the patient-exposed photon-flux or the cost of large area absorption gratings. The detailed system achieves a dark-field sensitivity adequate for lung imaging, governed by system autocorrelation lengths of up to 0.6 μm. The CT reconstructions show further evidence of the emergence of the dark-field in the parenchyma
Citation
Scientific Reports
Volume
15

Keywords

gold electrodeposition, gratings, imaging, superconformal, interferometer, lungs

Citation

Josell, D. , Spindler, S. , Pereira, A. , Organista, C. , Zdora, M. , Lovric, G. , Lautizi, G. , Dullin, C. , Contillo, A. , Longo, E. , Sodini, N. , D'Amico, L. , Romano, L. , Jefimovs, K. , Rawlik, M. , Tromba, G. and Stampanoni, M. (2025), Synchrotron radiation based analyzer-free dual-phase dark-field lung imaging approach for radiography and CT, Scientific Reports, [online], https://doi.org/10.1038/s41598-025-22355-x, https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=959280 (Accessed April 17, 2026)