Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Supplement 2 to the 2004 update of the AAPM Task Group No. 43 1 Report: Joint recommendations by the AAPM and GEC-ESTRO

Published

Author(s)

Michael G. Mitch, Mark J. Rivard, Facundo Ballester, Wayne M. Butler, Larry A. DeWerd, Geoffrey S. Ibbott, Christopher Melhus

Abstract

Since publication of the 2004 update to the American Association of Physicists in Medicine (AAPM) Task Group No. 43 Report (TG-43U1) and its 2007 supplement (TG-43U1S1), several new low- energy photon-emitting brachytherapy sources have become available. Many of these sources have satisfied the AAPM prerequisites for routine clinical purposes and are posted on the Brachytherapy Seed Registry managed jointly by the AAPM and the Imaging and Radiation Oncology Core Houston Quality Assurance Center (IROC Houston). Given increasingly closer interactions among physicists in North America and Europe, the AAPM and the Groupe Européen de Curiethérapie-European Society for Radiotherapy & Oncology (GEC-ESTRO) have prepared another supplement containing recommended brachytherapy dosimetry parameters for eleven low-energy photon-emitting brachytherapy sources. The current report presents consensus datasets approved by the AAPM and GEC-ESTRO. The following sources are included: 125I sources (BEBIG model I25.S17, BEBIG model I25.S17plus, BEBIG model I25.S18, Elekta model 130.002, Oncura model 9011, and Theragenics model AgX100); 103Pd sources (CivaTech Oncology model CS10, IBt model 1031L, IBt model 1032P, and IsoAid model IAPd-103A); and 131Cs (IsoRay Medical model CS-1 Rev2). Recommendations are presented on the selection of dosimetry parameters, such as from societal reports issuing consensus datasets (e.g., TG-43U1, AAPM Report #229), the joint AAPM/IROC Houston Registry, the GEC-ESTRO website, the Carleton University website, and those included in software releases from vendors of treatment planning systems. Depending upon the dosimetry parameters currently used by individual physicists, use of these recommended consensus datasets may result in changes to patient dose calculations. These changes should be carefully evaluated and reviewed with the radiation oncologist prior to their implementation.
Citation
Medical Physics
Volume
44

Keywords

TG-43, brachytherapy, dosimetry protocol, dosimetry parameters

Citation

Mitch, M. , Rivard, M. , Ballester, F. , Butler, W. , DeWerd, L. , Ibbott, G. and Melhus, C. (2017), Supplement 2 to the 2004 update of the AAPM Task Group No. 43 1 Report: Joint recommendations by the AAPM and GEC-ESTRO, Medical Physics (Accessed March 28, 2024)
Created September 30, 2017, Updated October 20, 2017