Anne-Marie Jeffery, John Q. Shields, Scott H. Shields
Abstract
A 4-terminal-pair ac bridge, capable of a relative uncertainty of 2 x 10^-9^, has been constructed at NIST. The bridge is a multi-ratio bridge (10:1, 2:1 and 1:1) and operates in the range from 100 Hz to 2000 Hz. The design and initial testing of this bridge is described.
Proceedings Title
Proc., Conference on Precision Electromagnetic Measurements (CPEM)
Jeffery, A.
, Shields, J.
and Shields, S.
(2000),
A Multi-Frequency 4-Terminal-Pair AC Bridge, Proc., Conference on Precision Electromagnetic Measurements (CPEM), Sydney, 1, AS, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=20272