Comparison of Two Josephson Array Voltage Standard Systems Using a Set of Zener References

Published

Author(s)

Yi-hua D. Tang, Richard L. Steiner, June E. Sims

Abstract

Two Josephson Array Voltage Standard Systems using different calibration algorithms are operated at the National Institute of Standards and Technology (NIST). A manual switch system is designed to make system comparisons by measuring against the 1.018 V of three Zener references. The difference of the average of the three Zener references is smaller than 1 nV.
Citation
IEEE Transactions on Instrumentation and Measurement
Volume
48
Issue
2

Keywords

Josephson voltage standard, calibration algorithms, system inter-comparison

Citation

Tang, Y. , Steiner, R. and Sims, J. (1999), Comparison of Two Josephson Array Voltage Standard Systems Using a Set of Zener References, IEEE Transactions on Instrumentation and Measurement, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=456 (Accessed April 23, 2024)