Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Acoustic measurement of the triple point of neon TNe and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry

Published

Author(s)

Changzhao Pan, Fernando Sparasci, Haiyang Zhang, Pascal Gambette, Mark Plimmer, Dario Imbraguglio, Roberto M. Gavioso, Michael R. Moldover, Bo Gao, Laurent Pitre

Abstract

We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt-Co resistance thermometer on the thermodynamic temperature scale by measuring the speed of sound in helium at a temperature T* chosen to be near the temperature of the triple point of neon, TNe. Prior to the present AGT, the Pt-Co thermometer was used with a neon triple-point cell as part of an interlaboratory comparison. Taken together, the results of the interlaboratory comparison and the present AGT redetermined the thermodynamic temperature TNe = (24.55515 ± 0.00024) K. This new value of TNe is consistent with other recent determinations obtained with various primary methods. After completing the AGT thermodynamic calibration, we used the Pt-Co thermometer to link T* to the temperature ratios measured by single-pressure refractive-index gas thermometry (SPRIGT) in a different laboratory. [Gao et al 2020 Metrologia 57 065006] Now, the T*-linked SPRIGT system can calibrate other thermometers on the thermodynamic temperature scale T in the range 5 K ≤ T ≤ TNe without using the International Temperature Scale ITS-90. At most temperatures in this range, the uncertainties of the T*-linked SPRIGT system are smaller than those of the ITS-90 systems used by National Metrology Institutes to calibrate resistance thermometers.
Citation
Metrologia
Volume
58

Keywords

thermometry, primary thermometry, Pt-Co thermometer, acoustic gas thermometry

Citation

Pan, C. , Sparasci, F. , Zhang, H. , Gambette, P. , Plimmer, M. , Imbraguglio, D. , Gavioso, R. , Moldover, M. , Gao, B. and Pitre, L. (2021), Acoustic measurement of the triple point of neon TNe and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry, Metrologia (Accessed April 18, 2024)
Created July 9, 2021, Updated February 24, 2023