Improved performance of the ECCS experiment at PTB
Published
Author(s)
Hansjoerg Scherer, Benedetta Camarota, Mark W. Keller, Sergey Lotkhov
Abstract
Recent improvements on the ECCS experiment at PTB are reported. A 5-junction single-electron R-pump is shown to shuttle electrons with a relative transfer error of 5×10-8. Moreover, the relative uncertainty of the setup for the cryogenic capacitor voltage measurement has been reduced to urel less than or equal to} 10-9. Further perspectives related to data evaluation and analysis on the ECCS experiment is also discussed.
Citation
IEEE Transactions on Instrumentation and Measurement
Scherer, H.
, Camarota, B.
, Keller, M.
and Lotkhov, S.
(2012),
Improved performance of the ECCS experiment at PTB, IEEE Transactions on Instrumentation and Measurement, [online], https://doi.org/10.1109/CPEM.2012.6250946