Performance of Multiplexed SQUID Readout for Cryogenic Sensor Arrays
Published
Author(s)
James A. Chervenak, Erich N. Grossman, Kent D. Irwin, John M. Martinis, Carl D. Reintsema, C. A. Allen, David I. Bergman, S. H. Mosely, Rick A. Shafer
Abstract
We report on the implementation of a multiplexer that users superconducting quantum interference devices (SQUIDs) to read out low-impedance cryogenic detectors. Using prototype chips, a circuit was built which interfaces eight input SQUID channels with a close-packed array of eight transition-edge sensor (TES) infrared bolometers. Circuit elements were measured and crosstalk specifications are reported. Digital feedback is employed to flux-lock a single element in the array of SQUIDs.
Citation
Nuclear Instruments & Methods in Physics Research Section A-Accelerators Spectrometers Detectors and Associated Equipment