Intercomparison of Visual Diffuse Transmission Density Measurements

Published

Author(s)

Egbert Buhr, Detleff Bergmann, E A. Early, Thomas R. O'Brian

Abstract

An intercomparison of visual diffuse transmission density measurements of photographic and x-ray film step tablets was completed bytwo national metrology institutes, the National Institute of Standards and Technology, USA, and the Physikalisch-Technische Bundesanstalt, Germany. The samples were measured by three reference densitometers according to procedures described in international documentary standards. The instruments are well characterized with estimates of their measurement uncertainties. The differences between the density measurements were less than 0.010 for the photographic films and 0.015 for the x-ray films fordensities as great as 3.3, with a systematic difference between the two institutes.
Citation
Journal of Imaging Science and Technology
Volume
44
Issue
No. 2

Keywords

densitometry, film step tablet, intercomparison, visual diffuse transmission density

Citation

Buhr, E. , Bergmann, D. , Early, E. and O'Brian, T. (2000), Intercomparison of Visual Diffuse Transmission Density Measurements, Journal of Imaging Science and Technology (Accessed March 29, 2024)