Reference 0 /45 Colorimeter

Published

Author(s)

Maria E. Nadal

Abstract

The Optical Technology Division at the National Institute of Standards and Technology (NIST) has developed a new instrument designed to measure the surface color of reflective, non-fluorescent samples at a geometry of 0 illumination angle and 45 viewing angle (0 /45 ). The instrument provides customers with means to incorporate calibrated transfer standards into their measurement systems and when required, a calibration service that can be used for measurements of customer supplied samples that meet the measurement criteria of the instrument. This document provides a detailed overview of the measurement methods, the uncertainty, and the characterization of the NIST Standard Reference 0/45 Colorimeter.
Citation
Special Publication (NIST SP) - 250-71
Report Number
250-71

Keywords

calibration, color standard, colorimeter, uncertainty

Citation

Nadal, M. (2008), Reference 0 /45 Colorimeter, Special Publication (NIST SP), National Institute of Standards and Technology, Gaithersburg, MD (Accessed March 28, 2024)