Author(s)
Jodie G. Pope, John D. Wright, Sherry D. Sheckels
Abstract
We report additional tests of our extended Lee model for calibrating turbine meters. The model accounts for 1) Reynolds number (Re) dependent drag and lift, 2) bearing static drag and 3) bearing viscous drag. Initially, we tested this model using a dual−rotor, 2.5−cm−diameter turbine meter and flow measurements spanning a 200:1 range (50 Re 109,000) with liquid mixtures spanning a 42:1 kinematic viscosity range (1.2 × 106 m2 / s ν 50 × 106 m2 / s). The model correlated the volumetric flow data within 3.6 % over the entire Re range. The same data had a maximum deviation of 17 % from the commonly used Strouhal versus Roshko (or Re) correlation. In this work, we tested the model using three different single−rotor turbine meters with diameters of 2.5 cm, 1.6 cm, and 1.9 cm and flow measurements spanning a 75:1 range (140
Conference Dates
June 20-22, 2012
Conference Location
Colorado Springs, CO
Conference Title
8th International Symposium on Fluid Flow Measurement
Keywords
Turbine Meter, Kinematic Viscosity, Propylene Glycol
Citation
Pope, J.
, Wright, J.
and Sheckels, S.
(2012),
Tests of the Extended Lee Model Using Three Different Turbine Meters, 8th International Symposium on Fluid Flow Measurement , Colorado Springs, CO, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=911318 (Accessed May 1, 2026)