Constant pressure primary flow standard for gas flows from 0.01 cm3/min to 100 cm3/min (10-8 mol/s to 10-5 mol/s)
Published
Author(s)
Robert F. Berg, Gooding Timothy, Robert E. Vest
Abstract
We describe a flow meter for gas flows in the range from 0.01 sccm to 100 sccm with a relative standard uncertainty of 0.03 % at 1 sccm. (1 sccm ≈ 1 cm3/min of an ideal gas at 101325 Pa and 0 C ≈ 0.7 mol/s.) The flow meter calibrates a secondary meter by withdrawing a piston from a cylinder held at constant pressure P while gas flows from the secondary meter into the cylinder. The flow meter can operate anywhere in the range 10 kPa
Berg, R.
, Timothy, G.
and Vest, R.
(2013),
Constant pressure primary flow standard for gas flows from 0.01 cm<sup>3</sup>/min to 100 cm<sup>3</sup>/min (10<sup>-8</sup> mol/s to 10<sup>-5</sup> mol/s), Flow Measurement and Instrumentation, [online], https://doi.org/10.1016/j.flowmeasinst.2013.12.002