Reflectance Factor Measurement Complications Due to Near Infrared Fluorescence

Published

Author(s)

Maria E. Nadal, Carl C. Miller, Clarence J. Zarobila

Abstract

The researchers at the National Institute of Standards and Technology (NIST) have found that the widely used orange and red color tiles unexpectedly exhibit near infrared fluorescence. An analysis of the effect this fluorescence component has on the measured reflectance factor utilizing different measurement methods.
Citation
Color Research and Application

Keywords

BCRA tiles, Color Standards, Fluorescence, Reflectance Factor

Citation

Nadal, M. , Miller, C. and Zarobila, C. (2010), Reflectance Factor Measurement Complications Due to Near Infrared Fluorescence, Color Research and Application, [online], https://doi.org/10.1002/col.20614 (Accessed March 28, 2024)
Created September 1, 2010, Updated November 10, 2018