Photodiode Calibration Comparison between Electro-Optic Sampling and Heterodyne Measurements up to 75 GHz
Published
Author(s)
Ari D. Feldman, Jeffrey A. Jargon, Tasshi Dennis, Paul D. Hale
Abstract
We present the first-ever comparison of a photodiodes measured frequency response calibrated with an electro-optic sampling system and a heterodyne system up to 75 GHz, along with the systems respective 95% confidence intervals. A brief description of each system and its known sources of uncertainties are provided. The two systems agree to within their respective uncertainties at most frequencies.
Proceedings Title
Conference on Precision Electromagnetic Measurements
Feldman, A.
, Jargon, J.
, Dennis, T.
and Hale, P.
(2020),
Photodiode Calibration Comparison between Electro-Optic Sampling and Heterodyne Measurements up to 75 GHz, Conference on Precision Electromagnetic Measurements, Denver, CO, [online], https://doi.org/10.1109/CPEM49742.2020.9191871