NGB 30m SANS - Small Angle Neutron Scattering

Perspective view of NG7 SANS instrument at NCNR

The 30 meter SANS instrument is installed on a split neutron guide NG-B. Designed to cover a wide Q-range, from 0.01 nm-1 using focusing refractive lenses to nearly 8.5 nm-1 at the lowest neutron wavelengths, it is suitable for examining structural features in materials ranging from roughly 1 to 700 nm. At least 75% of the instrument time will be allocated to guest researchers by the Beam Time Allocation Committee (BTAC).

Specifications/Capabilities

Q-Range0.01 nm-1 to 8.5 nm-1
Size Regime~1 nm to ~700 nm
SourceNeutron Guide (NG-B), cross-section: 60 mm x 60 mm
MonochromatorMechanical velocity selector with variable speed and pitch
Wavelength Range3.0 Å to 20.0 Å
Wavelength Resolution10% to 30% Δλ/λ (FWHM)
Source-to-Sample Distance4 m to 16 m in 1.5m steps via insertion of guide sections
CollimationCircular pinhole collimation or focusing biconcave lenses
Sample Size0 mm to 25 mm diameter
Sample-to-Detector Distance1.3 m to 13.1 m continuously variable
Detector640 mm x 640 mm 3He position-sensitive proportional counter with a 5.08 mm x 5.08 mm resolution
Optional CapabilitiesFocusing biconcave lenses
Power Outlets

220V, 30A, 1-phase - Qty. 2

208V, 30A, 3-phase - Qty. 2

208V, 60A, 3-phase - Qty. 2

110V, 30A, 1-phase - Numerous UPS and non-UPS

 

Scientific Opportunities/Applications

How NGB 30m SANS Beam Time is Typically Allocated in Each 38 Day Cycle:

  • 28 days minimum - BTAC Proposals
  • Balance (10 days or less) - Proprietary and Collaborative Research