Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

The Miami2001 Infrared Radiometer Calibration and Intercomparison. Part I: Laboratory Characterization of Blackbody Targets

Published

Author(s)

Joseph P. Rice, James J. Butler, B C. Johnson, P J. Minnett, K A. Maillet, T J. Nightingale, S J. Hook, A Abtahi, C J. Donlon, I J. Barton

Abstract

The second calibration and intercomparison of infrared radiometers (Maiami2001) was held at the University of Miami Rosenstiel School of Marine and Atmospheric Science (RSMAS) during May-June 2001. The participants were from several groups involved with the validation of skin sea-surface temperatures and land surface temperatures derived from the measurements of imaging radiometers on earth observation satellites. These satellite instruments include those on currently operational satellites and others that will be launched within two years following the workshop. There were two experimental campaigns carried out during the one week workshop : a set of measurements made by a variety of ship-based radiometers on board the research vessel F.G. walton-Smith in Gulf Stream waters off the eastern coast of Florida, and a set of laboratory measurements of typical external blackbodies used to calibrate these ship-based radiometers. this paper reports on the results obtained from the laboratory characterization of blackbody sources. A companion paper reports on the at-sea measurements.Five blackbody sources were intercompared by measurements of their brightness temperature using the National Institute of Standards and Technology (NIST) Thermal-infrared Transfer Radiometer (TXR). Four of these sources are used for calibration of sea-surface temperature radiometers. The fifth was a NIST water bath blackbody used for calibration of the TXR. All blackbodies agreed to better than 0.1 C at blackbody temperatures near the ambient room temperature. Some of the blackbodies had reduced effective emissivity relative to the NIST water bath blackbody, and hence they began to disagree at blackbody temperatures far enough away from the ambient room temperature. For these, relative effective emissivity values were determined so that corrections can be applied if they are used in conditions of non-laboratory ambient temperatures.
Citation
Journal of Atmospheric and Oceanic Technology
Volume
21
Issue
No. 2

Keywords

infrared, intercomparison, radiometry, remote, sea-surface, sensing temperature, thermal

Citation

Rice, J. , Butler, J. , Johnson, B. , Minnett, P. , Maillet, K. , Nightingale, T. , Hook, S. , Abtahi, A. , Donlon, C. and Barton, I. (2004), The Miami2001 Infrared Radiometer Calibration and Intercomparison. Part I: Laboratory Characterization of Blackbody Targets, Journal of Atmospheric and Oceanic Technology (Accessed March 29, 2024)
Created February 1, 2004, Updated February 17, 2017