Final Report of Key Comparison CCQM-K4 (Ethanol in Air)

Published

Author(s)

M J. Milton, C Brookes, A Marschal, Franklin R. Guenther, George C. Rhoderick, D E. Leer, L W. Zhen, C Takahashi, E Deak, Y A. Kustikov

Abstract

This key comparison involved primary standards of ethanol in air maintained at eight laboratories. The nominal amount fraction of the standards used for the comparison was 120 mmol/mol which is typical of the levels used to calibrate evidential breath analysers in many countries. Such standards fulfil the agreed OIML requirements for the calibration of evidential breath-alcohol analysers and are able to provide a more accurate calibration at field level than ethanol/water solution-based simulators which have a high sensitivity to variations in the temperature of the solution.
Citation
Metrologia

Keywords

CCQM, ethanol, gas standard, key comparison