Comparison of laser-based and monochromator-based thermodynamic temperature measurements

Published

Author(s)

George P. Eppeldauer, Howard W. Yoon, Jorge E. Neira, Vladimir Khromchenko, Charles E. Gibson, Allan W. Smith

Abstract

In this work, we describe the comparisons between the laser-based and monochromator-based radiance responsivity calibrations of a radiation thermometer. The spectral selection of the radiation thermometer is performed using a photopic-response filter which is spectrally broad in order to minimize the effects of convolution differences due to the spectral width of the laser- and the monochromator-based sources. The thick photopic-response filter also reduces the possible interference fringes which can be problematic in the laser-based calibrations. We will compare the radiance responsivities obtained using the two approaches and also compare the blackbody temperatures determined using the detector/monochromator based and the gold-point based calibration methods.
Proceedings Title
9th International Temperature Symposium
Volume
1552
Conference Dates
March 19-23, 2012
Conference Location
Anaheim, CA

Keywords

blackbody temperature measurement, kelvin, photopic filter, radiance responsivity, radiation thermometer, thermodynamic temperature measurement

Citation

Eppeldauer, G. , Yoon, H. , Neira, J. , Khromchenko, V. , Gibson, C. and Smith, A. (2012), Comparison of laser-based and monochromator-based thermodynamic temperature measurements, 9th International Temperature Symposium, Anaheim, CA, [online], https://doi.org/10.1063/1.4819516 (Accessed April 27, 2024)