NIST logo

Publication Citation: Absolute Cryogenic Radiometer and Solid State Trap Detectors for IR Power Scales Down to 1 pW with 0.1% Uncertainty

NIST Authors in Bold

Author(s): Adriaan C. Carter; Solomon I. Woods; Stephen Carr; Timothy M. Jung; Raju V. Datla;
Title: Absolute Cryogenic Radiometer and Solid State Trap Detectors for IR Power Scales Down to 1 pW with 0.1% Uncertainty
Published: June 02, 2009
Abstract: Commercially available Absolute Cryogenic Radiometers (ACRs) have combined uncertainties that grow rapidly above 1% at power levels below 10 nW. There are solid state detectors, however, used in sensors and radiometers that cannot be calibrated at levels as high as 10 nW because they begin to saturate. For this reason new detector-based standards are being developed to provide the low-background infrared calibration community with absolute traceability to powers down to 1 pW with ~0.1% combined uncertainty. The scale will be established using an ACR with a combined uncertainty on the order of 1 fW and a trap detector based on arsenic-doped silicon blocked impurity band devices with similar noise floor performance.
Proceedings: NEWRAD 2008
Volume: 46
Issue: 4
Pages: pp. S146 - S150
Location: Daejon, KR
Dates: October 13-16, 2008
Keywords: absolute cryogenic radiometer; blocked impurity band detector; IR radiometry; low power calibration; trap detector
Research Areas: Optical Properties of Materials
PDF version: PDF Document Click here to retrieve PDF version of paper (459KB)