Recertification of SRM 706a, a Polystyrene

Published

Author(s)

Kathleen M. Flynn, William R. Blair, J R. Maurey

Abstract

The recertification of the polystyrene standard reference material, SRM 706a, is described. The Mw of SRM 706a by light scattering was determined to be 2.85 x 105 g/mol with a sample standard deviation of 0.019 x 105 g/mol. A combined expanded uncertainty of 0.17 x 105 g/mol is estimated for this determination. The Mw obtained for on SRM 706 recently run. The Mw obtained for SRM 706a is also in excellent agreement with the Mw obtained by ultracentrifugation in 1967 of 2.88 x 105 g/mol on SRM 706. No bottle-to-bottle variability was found for SRM 706a in a size exclusion chromatography study.
Citation
NIST Interagency/Internal Report (NISTIR) - 6091
Report Number
6091

Keywords

average molecular weight, molecular weight standard, polystyrene standard, SRM 706, SRM 706a, Standard Reference Material, weight

Citation

Flynn, K. , Blair, W. and Maurey, J. (1998), Recertification of SRM 706a, a Polystyrene, NIST Interagency/Internal Report (NISTIR), National Institute of Standards and Technology, Gaithersburg, MD (Accessed April 16, 2024)
Created December 31, 1997, Updated October 12, 2021