NIST Authors in Bold
| Author(s): | Randall P. Wagner; Victor Nedzelnitsky; |
|---|---|
| Title: | Determination of Wave Motion Correction Values Required for Comparison Calibrations of a New Working Standard Microphone. |
| Published: | December 11, 2008 |
| Abstract: | Comparison calibrations of microphones based on the reciprocity method are performed with both the microphone under test and a reference standard microphone inserted into an acoustical coupler. For these calibrations, the wave motion correction W is needed to account for the departure from spatially uniform sound pressure amplitude in the coupler at frequencies for which the coupler interior dimensions are not small compared to a wavelength of sound. This internal report describes the procedure used to determine the values of W4192 applicable to comparison calibrations of a relatively new working standard microphone, the Bruel and Kjaer model 4192, and provides these values. |
| Citation: | NIST Interagency/Internal Report (NISTIR) - 7526 |
| Pages: | pp. 1 - 13 |
| Keywords: | Acoustical calibration; calibration of microphones; microphone calibration; NIST microphone calibration; reciprocity calibration; standard microphone; wave motion correction; working standard microphone. |
| Research Areas: | Electronics & Telecommunications |
| PDF version: | Click here to retrieve PDF version of paper (201KB) |