Abstract: Radiation force balance (RFB) measurements of time-averaged spatially-integrated ultrasound power transmitted into a reflectionless water load are based on measurements of the power received by the RFB target.  When conical targets are used to intercept the output of collimated circularly symmetric ultrasound sources operating at frequencies above a few megahertz, the correction for in-situ attenuation is significant, and differs significantly from predictions for idealized circumstances. Empirical attenuation correction factors for a 45degree (half-angle) absorptive conical NIST RFB target have been determined for 24 frequencies covering the 5 MHz-30MHz range.  They agree well with previously unpublished attenuation calibration factors determined in 1994 for a similar target.
Citation: Journal of Research (NIST JRES) -
Volume: 111
Pages: pp. 435 - 442
Keywords: attenuation correction;conical target;in-situ attenuation;power measurement;radiation force balance;radiation pressure;ultrasonic power;ultrasound power