NIST logo

Publication Citation: Error Compensation for CMM Touch Trigger Probes

NIST Authors in Bold

Author(s): William T. Estler; Steven D. Phillips; Bruce R. Borchardt; Ted Hopp; Christoph J. Witzgall; M Levenson; et al;
Title: Error Compensation for CMM Touch Trigger Probes
Published: January 01, 1995
Abstract: We present the analysis of a simple mechanical model of a common type of kinematic seat touch trigger probe widely used on modem coordinate measuring machines (CMMs). The model provides a quantitative description of the pre-travel variation or probe lobing characteristics that limit the use of such probes for high-accuracy dimensional measurements. We include the effects of stylus bending and the frictional interaction between the stylus ball and the part surface. The model is restricted to probes with simple straight styli, and we demonstrate significant error reduction both for vertically- and horizontally-oriented styli. In the latter case, gravitational forces are shown to play an important role in probe triggering and pre?travel variation. Extensions to arbitrary orientations are discussed.
Proceedings: Proceedings of ASPE Annual Meeting
Pages: pp. 428 - 431
Location: Austin, TX
Keywords: coordinate measuring machines;dimensional metrology;error correction;modeling;touch trigger probes
Research Areas: Metrology, Manufacturing