NIST Authors in Bold
| Author(s): | Gregory F. Strouse; |
|---|---|
| Title: | Standard Platinum Resistance Thermometer Calibrations from the Ar TP to the Ag FP |
| Published: | December 19, 2007 |
| Abstract: | The National Institute of Standards and Technology (NIST) is in charge of realizing, maintaining, and disseminating the international Scale of 1990 (ITS-90). The Standard Platinum Resistance Thermometer (SPRT) Calibration Laboratory of the NIST Thermometry Group realizes the ITS-90 from the argon triple point (Ar TP, 189.3442 °C) to the silver freezing point (Ag FP, 961.78 °C) for the calibration of SPRTs. This special publication describes the calibration services, methods, measurement assurance, and uncertainties for the NIST ITS-90 calibration SPRTs. |
| Citation: | NIST SP - |
| Keywords: | Fixed-point cells, International Temperature Scale of 1990, ITS-90, Standard Platinum Resistance Thermometer, SPRT, SPRT Cal. |
| Research Areas: | Thermometry |
| PDF version: | Click here to retrieve PDF version of paper (942KB) |