NIST logo

Publication Citation: A primary pressure standard at 100 kPa

NIST Authors in Bold

Author(s): James W. Schmidt; Y Q. Cen; Robert G. Driver; Walter J. Bowers Jr.; J C. Houck; S A. Tison; Charles D. Ehrlich;
Title: A primary pressure standard at 100 kPa
Published: June 01, 1999
Abstract: Primary pressure standards in the atmospheric pressure range are often established using mercury manometers. To a lesser extent controlled-clearance dead-weight testers in which one component (normally the piston) has been dimensionally measured have also been used. The recent advances of technology on two fronts 1) the fabrication of large diameter pistons and cylinders and 2) the dimensional metrology of those components have allowed some dead-weight testers at NIST to achieve total relative uncertainties in generated pressure near 10x10-6 (2ς). This paper described recent developments at NIST in which accurate dimensional measurements have been translated to effective areas. It is anticipated that total relative uncertainties in generated pressure may decrease to 5x10-6 (2ς) when recent dimensional measurements are incorporated on the newest gages.
Citation: Metrologia
Volume: 36
Issue: 6
Pages: pp. 525 - 529
Keywords: dead-weight testers;one atmosphere;pressure;primary;standards
Research Areas: Pressure & Vacuum, Chemistry