Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Uncertainty Budgets for Comparison Calibrations of Thermocouples

Published

Author(s)

Dean C. Ripple, K Garrity

Abstract

By means of concrete examples, we discuss the determination of an uncertainty budget for calibration of thermocouples by comparison techniques. In the first example, the calibration of a type T thermocouple (copper versus copper-nickel alloy) in a stirred liquid bath with a platinum resistance thermometer as a reference is considered. The second example is the calibration of a type K (a nickel-chromium versus nickel-aluminum alloy) thermocouple at temperatures up to 1200 C in a tube furnace, using a platinum-rhodium alloy thermocouple as a reference. Methods for determining the magnitudes of these uncertainties are given for each uncertainty component.
Proceedings Title
2005
Conference Dates
August 7-11, 2005
Conference Title
NCSL International Annual Workshop and Symposium

Keywords

calibration, thermocouple, thermometer, uncertainty

Citation

Ripple, D. and Garrity, K. (2005), Uncertainty Budgets for Comparison Calibrations of Thermocouples, 2005, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=830933 (Accessed April 19, 2024)