The Industrial Thermometer Calibration Laboratory calibrates
- Industrial thermometer probes
-
- platinum resistance (IPRT)
- thermistor
- thermocouple
- Digital readouts with attached probes
Calibrations in this laboratory are made by comparison with an ITS-90 calibrated standard platinum resistance thermometer in stirred liquid baths in the temperature range –196 ºC to 550 ºC. Fixed-point measurements at the Ga melting point (29.7646 °C) [for a thermometer with an outer diameter of less than 3.6 mm], the water triple point (0.01 °C) and the ice point (0 °C) are available.
How to Arrange a Calibration
- Contact the staff member in charge of the calibration service for your artifact with any technical or calibration cost questions.
- Calibrations are performed on a first come first serve basis.
- Ship your artifact with a copy of the purchase order. Use a sturdy box in which we can return the calibrated instrument.
- The purchase order may be sent to contact by e-mail, fax, or mail. Please include the following information:
- service ID for the calibration
- model and serial number of your artifact
- customer contact name, address, phone, and e-mail
- return shipping method (e.g., shipping account number)
| Mailing Address |
|
Shipping Address |
| NIST |
|
NIST |
| Attn: Dawn Cross |
|
Attn: Dawn Cross |
| 100 Bureau Drive |
|
100 Bureau Drive |
| MS 8363 |
|
Bldg. 221, Rm. B308 |
| Gaithersburg, MD 20899 |
|
Gaithersburg, MD 20899 |
Description and Cost of Services
|
Service ID
|
Fee/pt, US$
|
Description of Service
|
|
31110S
|
325
|
0 °C to 150 °C (32 °F to 300 °F )
|
|
31120S
|
441
|
151 °C to 315 °C (301 °F to 600 °F )
|
|
31130S
|
441
|
316 °C to 550 °C (601 °F to 1022 °F )
|
|
31140S
|
441
|
-1 °C to -110 °C (31 °F to -166 °F )
|
|
31150S
|
325
|
Liquid N2 (-196 °C or -321 °F )
|
Additional Services
|
Service ID
|
Fee, US$
|
Description of Service
|
|
31190S
|
278
|
Replacement report
|
|
31250S
|
117
|
Additional copy of report at time of test
|
|
31260S
|
at cost
|
Special thermometry services, by prearrangement
|
Calibration Measurement Uncertainties
Industrial Platinum Resistance Thermometers (IPRTs)
|
Temperature Range, °C
|
U (k=2), mK
|
|
-196
|
2.3
|
|
-97 to -80
|
9.0
|
|
-80 to -70
|
4.0
|
|
-70 to 0.5
|
2.3
|
|
0.5 to 95
|
2.4
|
|
95 to 275
|
4.8
|
|
275 to 550
|
7.5
|
Thermistors
|
Temperature Range, °C
|
U (k=2), mK
|
|
-196
|
1.8
|
|
-97 to -80
|
9.0
|
|
-80 to -70
|
3.6
|
|
-70 to 0.5
|
1.8
|
|
0.5 to 95
|
2.0
|
|
95 to 100
|
3.0
|
Thermocouples
|
Temperature Range, °C
|
U (k=2), °C
|
Thermocouple Type
|
|
-200 to 0
|
0.3
|
T
|
|
-200 to 0
|
0.2
|
E
|
|
-200 to 0
|
0.4
|
J
|
|
-200 to 0
|
0.4
|
K & N
|
|
0 to 500
|
0.5
|
K & N
|
Digital Readout with attached probes†
|
Temperature Range, °C
|
U (k=2), mK†
|
|
-196
|
2.3
|
|
-97 to -80
|
9.0
|
|
-80 to -70
|
4.0
|
|
-70 to 0.5
|
2.3
|
|
0.5 to 95
|
2.4
|
|
95 to 275
|
4.8
|
|
275 to 550
|
7.5
|
† The NIST calibration uncertainty does not include an uncertainty for the digital readout device.
General information regarding NIST calibration services may be found on the NIST Technology Services Calibration webpage.
|
|