Bookmark and Share Diffraction Metrology and Standards

Summary:

Our objective is the development of Standard Reference Materials (SRMs) and quantitative, reproducible, and accurate measurement methods for characterization of any structure possessing spatial order on the scale of X-ray wavelengths.  Such SRMs and measurement methods will aid in the development of new crystalline materials and devices made from them. Therefore, the impact of this project includes pharmaceuticals, ceramics, metals, semiconductors, and polymers.

Description:

Diffraction techniques can provide data on a number of samplePic 3 diffraction metrology and standards characteristics. Therefore, the method of certification and the artifact itself are chosen to address a specific measurement issue pertinent to a diffraction experiment.  NIST diffraction SRMs may be divided into five groups: Line Position, for calibration of the angle two-theta; Line Profile, for microstructure analysis; Instrument Response, for calibration of angle and intensity; Quantitative Analysis, for compositional analysis; and thin film SRMs for measurement of thickness, surface roughness, and density of layered structures. The most common use of NIST SRMs is for calibration of diffraction line position.  This requires that the SRM is certified with respect to lattice parameter.  This length must be measured in a manner that is traceable to the International System of Units (SI) meter through a robust and transparent measurement chain.  A major component of our effort consists of developing the capability for performing traceable measurement of lattice parameters for thin films and powders.

 

Impact and Customers

  • diffraction cline1Our primary impact is the enhancement of the measurement capability of the laboratory X-ray diffractometer; there are 20,000 in use worldwide.
  • Two of the three major instrument vendors supply NIST Diffraction Metrology SRMs with the equipment they sell. One vendor includes SRM 1976 with every instrument and, as a result, has been able to improve their angular resolution three times over the previous value.
  • Calibration of equipment with NIST Diffraction Metrology SRMs is requisite for ISO certification.
  • Primary customers are the International Center for Diffraction Data (ICDD) and X-ray instrument vendors (BEDE, Bruker AXS, Jordan Valley, PANalytical, Rigaku and Technos).

Major Accomplishments:

Work is continuous on the improvement of the measurement technology associated with the entire suite of diffraction metrology SRMs; three SRMs were certified this year. SRM 676a is the renewal of SRM 676, an alumina powder certified for quantitative analysis of multiphase mixtures. Improvements in SRM 676a include a certifiably pure-phase alumina powder, as well as a more accurate certification. SRM 1976a is the successor to SRM 1976, which, since its introduction in 1991, has become the best selling of the diffraction metrology SRMs. The new SRM 1976a features a more uniform microstructure that, in turn, results in reduced error bounds on the certified values. The new SRM 2000 is the first thin-film, high-resolution diffraction SRM. It consists of a 2.5 mm x 2.5 mm piece of a silicon wafer onto which a Si/Ge epilayer is grown. The SRM is certified with respect to the lattice parameter of the Si substrate.

While diffraction is commonly used for quantification of multiphase mixtures,using it to quantify the amorphous-component fraction of unknown samples has only become possible with the certification of  NIST SRMs for absolute phase purity. NIST's method for certification of phase content has been developed through interactions with the diffraction community over the last ten years.

Pic 4 analysis of diffraction data for certification
Analysis of Diffraction Data for Certification of SRM 676a

Our method's first application was in the recertification of SRM 676. In the new SRM 676a, further improvements to the experimental design realized the convergence of three measurement methods (neutron time-of-flight scattering and X-ray diffraction on two different synchrotron beamlines, at energies of 25 and 67 KeV); this allowed certification of SRM 676a with reduced error bounds. SRM 676a will also enable accurate measurement of the thickness of disordered surface layers, which is of particular importance for nanoscale materials with surface-dependent properties.

SRM 1976 was originally certified in response to a round robin study done by the International Centre for Diffraction Data in preparation for a new database format. A conclusion of the study was that a new SRM was needed that allowed for the calibration of both line position and intensity as a function of two-theta angle and instrument response using conventional data analysis methods. Based upon customer feedback on SRM 1976, the new SRM 1976a was manufactured in a single production run in order to significantly reduce the error bounds in the relative intensity data. SRM 2000 provides the high-resolution X-ray diffraction (HRXRD) community with either wavelength or angle calibration through certification of the substrate Si (220) lattice spacing. This SRM also provides certification of the Si wafer miscut (angle from surface to crystal plane) for sample rotation axis and sample holder alignment. Future recertification of this SRM will involve modeling HRXRD reflection data to provide customers with epitaxial layer structural information.

Recently, we presented the status of the powder diffraction suite of SRMs and progress on SRMs 676a, 1976a, and 2000 at the Denver X-ray Conference and the International Centre for Diffraction DataSpring Meeting.


Pic 5 SI traceability chart for SRM 2000

SI traceability chart for SRM 2000
diffraction_metrolog_image

Start Date:

October 24, 2008

End Date:

ongoing

Lead Organizational Unit:

MSEL