Report on Bilateral Key Comparison SIM.T-K6.1 on Humidity Standards in the Dew/Frost-point Temperature Range from –25 °C TO 20 °C

Published

Author(s)

Christopher W. Meyer, Ken Hill

Abstract

A Regional Metrology Organization (RMO) Key Comparison of dew/frost point temperatures was carried out by the National Institute of Standards and Technology (NIST, USA) and the National Research Council (NRC, Canada) between December, 2014 and April, 2015. The results of this comparison are reported here, along with descriptions of the humidity laboratory standards for NIST and NRC and the uncertainty budget for these standards. This report also describes the protocol for the comparison and presents the data acquired. The results are analyzed, determining degree of equivalence between the dew/frost-point standards of NIST and NRC.
Citation
Metrologia
Volume
52

Keywords

Key Comparison, Humidity, Dew Point, Frost Point, Degree of Equivalence.

Citation

Meyer, C. and Hill, K. (2015), Report on Bilateral Key Comparison SIM.T-K6.1 on Humidity Standards in the Dew/Frost-point Temperature Range from –25 °C TO 20 °C, Metrologia, [online], https://doi.org/10.1088/0026-1394/52/1A/03008 (Accessed April 19, 2024)
Created September 30, 2015, Updated April 14, 2022