Publication Citation: Test Report-Exposure and Ambient dose Equivalent Rate Measurements in Support of the ITRAP+10 Testing

NIST Authors in Bold

Author(s): Leticia S. Pibida; Ronaldo Minniti; Larry L. Lucas; C M. O'Brien;
Title: Test Report-Exposure and Ambient dose Equivalent Rate Measurements in Support of the ITRAP+10 Testing
Published: August 01, 2013
Abstract: In this work we studied the response of two different Victoreen® instruments as a function of the exposure rate, the instrument orientation and photon energy. The rate dependence for both instruments is of the order of 8 % over the range of exposure rates tested (0.5 mR/h to 1000 mR/h). Regarding the instrument orientation dependence investigation, a significant difference is observed between the two instruments. While the Victoreen® 451P shows no significant dependence with instrument orientation, the Victoreen® 451P-DE-SI shows a significant dependence of up to 20 % between the three different orientations. Finally, the energy dependence measurements of both instruments reveal the instruments measure lower exposure rate values compared to the reference values for the low energy x-rays by about 20 %. While, for the 60Co gamma-ray beam measurements, both instruments are about about 5 % higher than the reference value.
Citation: NIST TN - 1800
Keywords: exposure rate; ambient dose equivalent rate; ITRAP+10 testing; Victoreen 451P; Victoreen 451P-DE-SI
Research Areas: Radiation Physics, Homeland Security and Disaster Resilience