Characterization of Two Spectrmeters in Support of the Landsat Data Continuity Mission

Published

Author(s)

Bettye C. Johnson, Robert D. Saunders, Zhigang Li, Abra Fein, Lawrence Ong, Milton Hom, Robert Barnes, B L. Markham

Abstract

The Landsat Data Continuity Mission (LDCM) project at the National Aeronautics and Space Administration (NASA)Goddard Space Flight Center (GSFC) is supervising the manufacture and calibration of the Operational Land Imager (OLI) satellite instrument by Ball Aerospace in Boulder, Colorado. As part of that oversight function, the project is preparing a set of radiometers to monitor long-term changes (if any) in the radiance from the integrating sphere used for the radiance calibration of the OLI instrument. That sphere, calibrated at the National Institute of Standards and Technology (NIST), serves as an artifact for establishing traceability of the OLI radiance calibration to SI units, that is, to the radiance scale at NIST. This paper addresses the characterization of two Analytic Spectral Devices (ASD) Fieldspec spectrometers that are part of the NASA/NIST program to validate radiometric reference standards in the LDCM project. In particular, we report on a series of measurements at NIST to determine the ASD spectrometers long-term stability. Along with other radiometers, the ASDs will be used in the monitoring of changes in the OLI reference sphere from its calibration at NIST to its use in the calibration of the OLI satellite instrument. The ASD stability measurements will continue through the conclusion of the calibration of OLI.
Proceedings Title
SPIE Optical Engineering + Applications
Conference Dates
August 1-5, 2010
Conference Location
San Diego, CA
Conference Title
Proc. SPIE 7807

Keywords

LDCM, OLI, Radiance, Prelaunch Calibration

Citation

Johnson, B. , Saunders, R. , Li, Z. , Fein, A. , Ong, L. , Hom, M. , Barnes, R. and Markham, B. (2010), Characterization of Two Spectrmeters in Support of the Landsat Data Continuity Mission, SPIE Optical Engineering + Applications , San Diego, CA, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=906346 (Accessed March 29, 2024)