NIST logo

Publication Citation: Comparison of Results of Calibrating the Magnitude of the Sensitivity of Accelerometers by Laser Interferometry and Reciprocity

NIST Authors in Bold

Author(s): Beverly F. Payne; David J. Evans;
Title: Comparison of Results of Calibrating the Magnitude of the Sensitivity of Accelerometers by Laser Interferometry and Reciprocity
Published: January 01, 1999
Abstract: Minimizing the uncertainty in the calibration of accelerometers by the reciprocity method requires minimizing the cross-axis component of the induced motion, and the distortion of its waveform. Minimizing the uncertainty in the calibration of accelerometers by the laser interferometric fringe-counting method requires minimizing the mechanical coupling between the moving element of the shaker and the optical components of the interferometer. The design of the new shaker provides for minimal distortion and cross-axis motion in combination with a very large mechanical impedance between the moving element of the shaker and the optical components of the interferometer. The shaker is equipped with dual coils and two retractable magnets to provide for reciprocity measurements without having to attach a secondary source of vibration located external to the moving element. This paper presents a comparison of calibration results obtained using the two methods and describes the shaker and calibration systems.
Citation: Metrologia
Volume: 36
Issue: (4)
Pages: pp. 391 - 394
Keywords: accelorometers;calibration;exciters;fringe counting;laser interferometer;reciprocity shakers;uncertainty
Research Areas: Metrology