Publication Citation: Secondary Ion Mass Spectrometry Measurements of Isotopic Ratios: Correction for Time Varying Count Rate

NIST Authors in Bold

Author(s): Kevin J. Coakley; A M. Leifer; David S. Simons;
Title: Secondary Ion Mass Spectrometry Measurements of Isotopic Ratios: Correction for Time Varying Count Rate
Published: January 01, 2005
Abstract: In Secondary Ion Mass Spectrometry measurement systems, the count rate of isotopes may vary in time as a particle is consumed during the analysis. Since only one isotope at a time is measured, this drift can introduce systematic error into the estimate of the ratio of any two isotopes. We correct the measurements for drift by aligning the time series of isotopic
Citation: The International Journal of Mass Spectrometry
Volume: 240
Pages: pp. 107 - 120