1 Volt Josephson Fast Reversed DC Source

Published

Author(s)

Charles J. Burroughs, Samuel Benz, Todd E. Harvey, H. Sasaki

Abstract

We have made several improvements to the NIST Josephson FRDC (fast reversed dc) source so that it now operates at 1 Volt. We have made a precision comparison to a conventional semiconductor FRDC source by measuring the thermoelectric transfer difference of a known thermal voltage converter. The agreement between the two sources was measured to be better than 1 part in 10^u7.
Proceedings Title
Proc., Conference on Precision Electromagnetic Measurements (CPEM)
Issue
22
Conference Dates
May 14-19, 2000
Conference Location
Sydney, 1, AS

Keywords

digitaal-to-analog converter, Josephson array, metrology, standard, voltage

Citation

Burroughs, C. , Benz, S. , Harvey, T. and Sasaki, H. (2000), 1 Volt Josephson Fast Reversed DC Source, Proc., Conference on Precision Electromagnetic Measurements (CPEM), Sydney, 1, AS, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=10539 (Accessed March 29, 2024)