12/12/2003
|
2/4/2003
Given one Energy Dispersive Spectrum (EDS) peak width (principal peak),
for a Si(Li) detector
the widths of the other peaks are
sqrt[2500(Ei - Ep) + (2355 Sp)^2]
Si = ---------------------------------
2.355
Si standard deviation of peak i, keV
Sp standard deviation of the principal peak, keV
Ei energy of peak i, keV (I think).
Ep energy of principal peak
from p. 386, eq. 7.19
Goldstein, J.I., et. al.
SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS, 2nd Ed.
Plenum, N.Y. 1992
Original work:
Fiori, C.E., Newbury, D.E., (1978), SEM 1:401.
------------------------------------------------------------------
2/4/2003 The above may have dimension problems. Dale will check it out.
In the mean time, a simplified equation from the original article is:
FWHMx^2 = 2.5(Ex - Eref) + FWHMref^2
Where all units are eV.
Mn usually principal peak 5895 eV 151 ev
Al calculated from formula 1498 eV 107 ev
maclispix(45): (line-fwhm 1498 5895 151)
108.66692
close enough.
---------------------------------------------------------------------
2/28/2003 Trouble:
Lispix(38): (xray-line-fwhm 1600 6392 96)
#c(0.0 52.573757)
Trouble point: 2.5(Ex - Eref) + FWHMref^2 =< 0
or: 2.5(Ex - Er) + Wr^2 <= 0
2.5(Ex - Er) <= -Wr^2
2.5(Er - Ex) >= Wr^2
Edif >= Wr^2/2.5
4800 * 2.5 = 12000 - sqrt = 109.
Ex = 1600
Er = 6392
sqrt (2.5 (Er - Ex)) = 109.4
Lispix(47): (xray-line-fwhm 1600 6392 120)
49.193497
Lispix(48): (xray-line-fwhm 1600 6392 110)
10.954452
???
Ex - Eref >= FWHMref^2 / 2.5
Also, the Si peak range looks really broad - small side peak adds to the width. That ain't right.